Traditional G-JET is engineered with an innovative internal tube design that transitions from a triple-lumen configuration proximally to a single-lumen structure distally, specifically developed to help minimize clogging and support consistent enteral feeding performance. This advanced geometry improves flow dynamics while reducing areas where formula residue may accumulate. Manufactured from medical-grade silicone, the Traditional G-JET is designed to enhance patient comfort while maintaining durability and long-term reliability. Its thoughtful construction supports safe placement, secure gastric sealing, and improved jejunal feeding efficiency when used under appropriate medical supervision.
Features
- Improve Comfort: Smooth silicone construction and refined internal geometry are designed to enhance comfort and tolerance at the stoma site.
- AMT Exclusive Balloon: Apple-shaped balloon expands horizontally to create a more effective gastric seal, helping reduce leakage at the stoma site.
- Patient Comfort: Constructed from soft, medical-grade silicone to help minimize skin irritation while enhancing overall patient comfort during long-term use.
- Reduce Clogging: Proprietary internal design transitions from a triple lumen to a single lumen to support improved flow and help reduce the likelihood of clogging.
- Anti-Kink Technology: Integrated spring reinforcement helps stabilize the jejunal lumen to reduce kinking and support uninterrupted flow; device is MR Conditional (*available in 16F & 18F sizes only).
The Traditional G-JET kit includes: devices (1), small syringe (1), large syringe (1), gauze.
The Traditional G-JET can accommodate gastric stoma lengths up to 10 cm.
ENFit is a registered trademark of GEDSA, Inc. or its affiliates.
NOTE: G-JET devices should never be cut to custom lengths.
The Traditional G-JET is cleared for initial placement and is MR Conditional.
NOTE: The Traditional Length G-JET may be placed percutaneously under fluoroscopic or endoscopic guidance or as a replacement to an existing device using an established stoma tract. Placement/Removal is to be performed by a qualified clinician.
Balloon Fill Volumes
| French Size |
Minimum |
Recommended |
Maximum |
| 14F |
3 ml |
4 ml |
5 ml |
| 16F |
3 ml |
4 ml |
5 ml |
| 18F |
7 ml |
8 ml |
9 ml |
| 22F |
8 ml |
10 ml |
12 ml |
Note: Recommended fill volume is clearly printed on the balloon fill valve of the Traditional Length G-JET. Consult your healthcare provider before adjusting fill volumes.