The Omnis Health Embrace TALK HI Control Solution is an essential tool for ensuring the proper functioning of the Embrace TALK Blood Glucose Monitor and Embrace GlucoseTest Strips. Control solution testing should be performed under various circumstances to verify the accuracy and reliability of glucose readings.

Features

  • Compatible exclusively with Embrace Talk Blood Glucose Monitor for dependable performance.
  • Easy application with one drop on the test strip replaces a blood sample for a control test.
  • Ensures the glucose meter and test strips are working together properly.
  • Comes in a 2.5 mL volume, sufficient for multiple testing sessions.
  • Control range validation accessible on the side of the vial of test strips for user convenience.
     

When to Perform Control Solution Testing:

  • When using the monitor for the first time
  • When using a new bottle of Embrace TALK Blood Glucose Test Strips
  • After leaving the test strip bottle cap open for an extended period
  • Following an incident where the monitor may have been damaged, such as dropping it
  • If there are suspicions that the monitor or test strips are not functioning correctly
  • When blood glucose test results do not align with how the user feels
     

Performing a Control Solution Test:

Step 1: Preparation

  • Check the expiration date of the control solution and ensure it is not expired.
  • Determine whether the control solution is labeled as LOW or HIGH and set the monitor accordingly (L1 for LOW, L2 for HIGH).
  • Gently mix the control solution by rolling the bottle between your hands.
  • Place 2 drops of control solution on a clean glass or plastic plate and immediately replace the cap on the control solution bottle.
     

Step 2: Insert Test Strip
 

  • Insert an Embrace TALK Test Strip into the monitor and close the lid promptly.

Step 3: Set Monitor

  • Turn on the monitor by inserting the test strip. The monitor will prompt with audio instructions if the talking feature is enabled.
  • Press the appropriate button based on the level of control solution used (bottom button for LOW-L1, top button for HIGH-L2). The monitor will display L1/L2 accordingly.
     

Step 4: Apply Control Solution

  • Gently guide the tip of the test strip into the drop of control solution. The monitor will play a melody and begin a countdown.
  • After 6 seconds, the control solution reading will be displayed on the monitor.
     

Step 5: Compare Results

  • Compare the displayed reading with the control solution range indicated on the bottle of test strips.
  • If the reading falls within the specified range for the corresponding level (L1 or L2), the monitor and strips are functioning correctly.
  • Follow these steps for both Level 1 and Level 2 Control Solution tests to ensure the accuracy of the Embrace TALK Blood Glucose Monitor and Test Strips.
     

Glucose Control Solution Troubleshooting:
 

Possible Causes What you can do
Wrong brand of Control Solution being used Make sure you are using Embrace TALK Control Solution.
Control Solution not at room temperature Make sure the testing environment is between 50�F to 104�F, <85% RH.
Expired or contaminated control solution Check the expiration and open bottle date on both the control solution and test strips. (Test strips should be used within 6 months from date opened. Control solution should be used within 3 months from date opened). Repeat testing using a new test strip. If the result is still out of range, use a new bottle of control solution and retest.
Damaged test strip
Monitor malfunction If problem persists, call Omnis Health@1-877-979-5454.
 
Warranty
  • The product warranty is applicable as per the terms and conditions provided by the product manufacturer.

Please call us for specific details.

Return policy
  • No returns will be accepted after 30 days from the date of shipment.
  • All returns are subject to a restocking fee as per manufacturers terms and conditions.
  • All returns must have an RGA number (Returned Goods Authorization), unauthorized returns will not be accepted.
  • We do not guarantee fulfillment of any desired purpose or product suitability to the user and this will not be considered as a valid reason for return.
  • The products must be new, unused condition, not tampered with, in original packaging and returned at the customers expense in the original packaging.
  • If your return is not due to any manufacturing defect then the original shipping cost will be deducted from the total refund.
  • Hygiene, bath and toilet items cannot be returned once opened or used.
  • Standard manufacturer terms and conditions apply for return policy of this product.

Please call us for specific details.

Looking for more information?

Select Variant
Embrace TALK HI Control Solution
$2.15$2.26
Select Choice :
  • High Flow

  • Low Flow

  • Select Group